FLUKE 572-2 High Temperature Infrared Thermometer

The Fluke 572-2 High-Temperature Infrared Thermometer is the ideal tool for high-temperature industrial environments all around the world. Whether you work in power utility, metal refining and smelting, glass, cement or petrochemical environments, the 572-2 offers the rugged performance accuracy you need to get the job done, backed by the most trusted name in test tools. The simple, three-button on-screen menu interface saves time and makes even complex measurements easy. With just a few pushes of a button you can adjust emissivity, start data logging, or turn on and off alarms. The 60:1 distance-to-spot ratio with dual laser sighting helps pinpoint the target fast, making it easier to measure small objects from a long distance.

Key Features :

• High temperature infrared thermometer that measures from -30°C to 900°C (-22°F to 1652°F)
• Offers an ultra-high 60:1 distance-to-spot ratio with dual laser sighting for fast, accurate targeting
• Features a user-selectable multi-language interface
• Displays the temperature plus MAX, MIN, DIF, AVG temperature
• Provides adjustable emissivity and a predefined emissivity table
• Audible and visible alarms for rapid detection of high and low temperatures outside the limits
• Stores up to 99 data sets for review and analysis
• Powered by two, standard AA batteries -30-
• Presents infrared and thermocouple temperatures on a bright backlit display
• Is compatible with standard mini-connector K-type thermocouples, including ones you already own and have installed
• Features last reading Hold (20 seconds)

Specifications :

Infrared Measurements
Infrared temperature range-30°C to 900°C (-22°F to 1652°F)
IR accuracy (calibration geometry with ambient temperature 23°C ± 2°C)≥ 0°C: ± 1°C or ± 1% of the reading, whichever is greater
≥ -10°C to < 0°C: ±2°C
< -10°C: ±3°C
≥ 32°F: ± 2°F or ± 1% of the reading, whichever is greater
≥ 14°F to < 32°F: ±4°F
< 14°F: ±6°F
IR repeatability±0.5% of reading or ±0.5°C (±1°F), whichever is greater
Display resolution0.1°C / 0.1°F
Distance: Spot60:1 (calculated at 90% energy)
Minimum spot size19 mm
Laser sightingOffset dual laser, output < 1 mW
Spectral response8 μm to 14 μm
Response time (95%)< 500 ms
EmissivityDigitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials
Contact Measurements
K-type thermocouple input temperature range-270°C to 1372°C (-454°F to 2501°F)
K-type thermocouple input accuracy (with ambient temperature 23°C ± 2°C)< -40°C: ±(1°C + 0.2° / 1°C)
≥ -40°C: ±1% or 1°C, whichever is greater
< -40°F: ±(2°F + 0.2° / 1°F)
≥ -40°F: ±1% or 2°F, whichever is greater
K-type thermocouple0.1°C / 0.1°F
K-type thermocouple repeatability±0.5% of reading or ±0.5°C (±1°F), whichever is greater

 